Publicaciones científicas

Quantitative assessment of left ventricular size and function in cardiac transplant recipients: Side-by-side comparison of real time two-dimensional echocardiography, contrast-enhanced two-dimensional echocardiography, three-dimensional echocardiography, and contrast-enhanced three-dimensional echocardiography as compared to magnetic resonance imaging

01-feb-2019 | Revista: Echocardiography

Moises Rodriguez-Mañero  1   2   3 , Pedro Azcárate-Agüero  4 , Bahij Kreidieh  5 , María Teresa Alvez  1 , Amparo Martínez-Monzonís  1   2   3 , Agnes Diaz-Dorronsoro  6 , Adrian Cid-Menéndez  1 , José Ramón González-Juanatey  1   2   3 , Joaquin Barba-Cosials  6 , Gregorio Rábago  6 , Gorka Bastarrika  6


Introduction: We evaluate the ability of 2D non-contrast-enhanced echocardiography (CE-echo), 2DCE-echo, 3D-echo, 3D non-CE-echo, and 3DCE-echo to evaluate allograft function and dimensions in orthotropic heart transplantation (OHT). Cardiac resonance (CMR) was used as reference.

Methods: Twenty six consecutive OHT-recipients were prospectively recruited. Bland-Altman, Spearman rank, and concordance-correlation coefficients (CCC) were determined.

Results: Good CCCs were found between the four modalities and CMR for ejection fraction (r ≥ 0.72/P < 0.001; r ≥ 0.77/ P < 0.001; r ≥ 0.51/ P < 0.23; r ≥ 0.75/ P < 0.001, respectively). Highest intraclass correlation coefficient (ICC) was for 2D CE-echo(CCC = 0.77). End-diastolic volume(EDV) measurements statistically differed when 2D non-CE-echo, 2DCE-echo, and 3D non-CE-echo were compared with the cross-sectional imaging modalities, but they did not differ significantly from 3DCE-echo. End-systolic volume (ESV) and stroke volume (SV) differed statistically between the four modalities; however, SV measured by CMR and 3DCE-echo were comparable. Overall, 2D non-CE-echo, 2DCE-echo, and 3D non-CE-echo showed lower mean EDV, ESV, and SV than CMR. ICC was that of the ESV variable in the 4 techniques, with the values of the ICC of the 3DCE-echo technique superior to the rest. Overall, the best CCC were found for 3DCE(r = 0.88, 0.92 and 0.76 for EDV, ESV and SV, respectively).

Conclusion: Routine use of 3DCE-echo may allow more comprehensive cardiac assessment in cardiac transplant recipients.

CITA DEL ARTÍCULO  Echocardiography. 2019 Feb;36(2):306-311. doi: 10.1111/echo.14232. Epub 2019 Jan 1.